SubjectRe: (subset) [PATCHv2 0/5] Support for GE B1x5v2 and B1x5Pv2
Date
On Thu, 29 Apr 2021 00:29:48 +0200, Sebastian Reichel wrote:
> This series adds support for another General Electric patient
> monitor series (similar to existing Bx50v3), which is based on
> i.MX6DL using Congatec's QMX6 module.
>
> The module uses an I2C RTC to provide the i.MX6 32768 Hz clock,
> so it's important to keep it enabled. Not doing so results in
> incorrect timings of watchdog and i.MX6 RTC. The bootloader
> enables the watchdog, so disabling the clock results in system
> reboot. [0]
>
> [...]

Applied, thanks!

[1/5] rtc: m41t80: add support for fixed clock
commit: f765e349c3e1f2e676ad4bd61197216b26976022

Best regards,
--